DT&C Vina — A Pioneer in Rigorous Compliance with SAR Standard QCVN 134:2024/BTTTT

In the era of wireless connectivity, assessing the Specific Absorption Rate (SAR) is not merely a regulatory passport for electronics and telecommunications products to enter the most demanding markets — it is a fundamental measure of user safety. An inaccurate measurement result can lead to serious consequences: either overestimating risk (causing economic loss to the business) or underestimating it (posing health hazards to end users).

Effective 15 February 2025, the National Technical Regulation QCVN 134:2024/BTTTT on Specific Absorption Rate (SAR) for hand-held and body-worn wireless communication devices officially came into force in Vietnam. This regulation establishes the mandatory legal framework for testing, conformity certification, and declaration of conformity for mobile phones, laptops, tablets, and DECT wireless telephone devices before they can be placed on the Vietnamese market.

QCVN 134:2024/BTTTT sets out numerous stringent technical requirements. In addition to requirements on test systems and procedures, the testing environment conditions are also very strictly regulated — forming the essential foundation for SAR measurements that are accurate, reliable, and legally valid. Join Dt&C Vina as we explore the stringent technical requirements for the SAR testing environment in this article.

 

I. QCVN 134:2024/BTTTT — Vietnam's National Technical Regulation on SAR

QCVN 134:2024/BTTTT directly references the international standard IEC/IEEE 62209-1528:2020 as the technical basis for all SAR measurement methods. This means that all requirements regarding test equipment, measurement environment, test procedures, and measurement uncertainty set out in IEC/IEEE 62209-1528:2020 become mandatory requirements when conducting SAR testing under QCVN 134:2024/BTTTT in Vietnam.

 

II. Stringent Technical Requirements for the SAR Testing Environment

Since QCVN 134:2024/BTTTT references IEC/IEEE 62209-1528:2020, the testing environment requirements set out in this international standard become mandatory compliance obligations in Vietnam.

The two most critical requirement groups are:

 

2.1 Temperature Control Requirements

       The ambient temperature of the measurement room and the temperature of the tissue-equivalent liquid must be maintained within the range of 18°C to 25°C.

       The temperature of the tissue-equivalent medium during SAR measurements must remain within ±2°C of the temperature at which the dielectric properties were measured. If the temperature deviation exceeds this limit, the dielectric properties must be re-measured.

       Dielectric properties must be measured within 24 hours prior to conducting SAR measurements and must be re-measured after every 48 hours of continuous use.

 

2.2 Ambient Electromagnetic Interference Control Requirements

       The SAR error caused by ambient electromagnetic interference from the surrounding environment must be ≤ 0.012 W/kg (equivalent to 3% of the reference level of 0.4 W/kg).

       If this threshold is exceeded, the test laboratory must either apply mitigation measures or demonstrate that the interference contribution is less than 10% of the actual SAR value of the device under test, and account for it in the measurement uncertainty budget.

       The device under test must not be connected to any live wireless network during the measurement.

 

III. Dt&C Vina — Achieving the Highest Level of Compliance with QCVN 134:2024/BTTTT

Dt&C Vina has proactively chosen the highest level of environmental control — placing each SAR measurement system inside a dedicated, independent shielded chamber (shield room).


 

 

3.1 Independent Shielded Chamber System

With dedicated independent shielded chambers (shield rooms) for each SAR measurement system, Dt&C Vina ensures:

       Ambient electromagnetic interference levels are maintained at near-zero — significantly below the 0.012 W/kg threshold required by the standard, completely eliminating measurement errors caused by environmental interference.

       Each chamber operates fully independently — with no cross-interference, enabling simultaneous testing across multiple projects.

 

3.2 Dedicated Precision Climate Control Systems — Maintaining Stable Temperature 24/7 Across All SAR Testing Rooms.

The temperature control system in each shielded chamber is independently configured and managed, maintaining stable conditions within the 18°C–25°C range as required by QCVN 134:2024/BTTTT and the referenced standard IEC/IEEE 62209-1528:2020.


 


 

Dt&C Vina has installed a dedicated air conditioning system for the SAR test chambers

 

 

IV. Practical Benefits for Customers

 

       Parallel Testing Across Multiple Projects: Independent shielded chambers allow simultaneous testing of multiple product projects at the same time.

       Accurate and Reproducible Results: A fully stable and controlled measurement environment eliminates random errors caused by environmental interference. Dt&C Vina's measurement results offer high reproducibility, suitable for both repeat testing and longitudinal result comparison.

       Reliable Turnaround Commitments: With no dependency on external wireless conditions, Dt&C Vina faces no risk of measurement delays caused by unexpected interference. Customers receive firm testing schedules and guaranteed result delivery timelines.

       Multi-Market Report Validity: Test reports issued by Dt&C Vina fully satisfy the requirements of QCVN 134:2024/BTTTT for the Vietnamese market, while also serving as a strong reference for SAR-regulated export markets including FCC (USA), CE (EU), MIC (Japan), and ISED (Canada).

       Product Information Confidentiality: Each project is conducted within a completely physically isolated chamber, ensuring that project information and test samples remain strictly confidential, in full compliance with non-disclosure agreements signed with customers.

V. Conclusion

QCVN 134:2024/BTTTT marks a pivotal milestone in the quality management of wireless devices in Vietnam. Dt&C Vina is committed to continuously investing in world-class infrastructure and equipment, and to rigorously adhering to testing procedures — ensuring customers receive the highest quality SAR testing services: accurate, reliable, and ready to meet the demands of both domestic and international markets.